Thickness of the cornea and the corneal flap in Laser In Situ Keratomileusis measured by an ultrasound

潘红飙,吴章友,顾永昊,温耀春
DOI: https://doi.org/10.3969/j.issn.1000-0399.2009.06.12
2009-01-01
Abstract:Objective To investigate the thickness of the cornea and the corneal flap in laser in situ keratomileusis (LASIK) measured by an ultrasound. Methods Fifty five (55) cases (110 eyes) of myopia received LASIK. Moria M2 microkeratome was used to make the corneal flaps. At location of these centers, 3:00, 6:00, 9:00, and 12:00, the thickness of cornea and residual corneal substance were measured by an ultrasound before and after the surgery. The thickness of the corneal flap was calculated by subtracting the residual corneal stromol thickness from the thickness of cornea. Resealts The corneal flap thickness at different locations made by Moria M2 microkeratome was in equable; Actual thickness of corneal flap was different from expected value; There was obvious difference of thickness between two corneal flap made by same microkeratome. Conclusion An ultrasound could provide valuable data for LASIK and make the surgery more safe, stable and predictable.