Calibration of depth-coordinate of spectral domain OCT image

Lin Ling,Li Gang,Zheng Yu
DOI: https://doi.org/10.19650/j.cnki.cjsi.2008.03.004
2008-01-01
Abstract:To calibrate the depth-coordinate of Spectral Domain OCT (SDOCT) image, a method was suggested and tested with sample images. The depth-coordinate interval of a SDOCT image is not equidistant because it is determined by the characteristics of the light source and the refractive index of the anisotropic sample. The maxim depth-coordinate (the spatial range of a SDOCT) depends on the total pixel number of the liner CCD and the virtual position of the reference mirror: the maximum depth-coordinate limited by the wavelength resolution of the spectrometer is proportional to the total number of pixels. When the virtual position of the reference mirror is inside the sample, the maximum depth-coordinate is double of that when it is outside the sample. This method can be applied in Spectral Domain OCT.