Full-range space-division multiplexing optical coherence tomography angiography
Yongyang Huang,Jason Jerwick,Guoyan Liu,Chao Zhou
DOI: https://doi.org/10.1364/boe.400162
IF: 3.4
2020-07-31
Biomedical Optics Express
Abstract:In this study, we demonstrated a full-range space-division multiplexing optical coherence tomography (FR-SDM-OCT) system. Utilizing the galvanometer-based phase modulation full-range technique, the total imaging range of FR-SDM-OCT can be extended to >20 mm in tissue, with a digitizer sampling rate of 500 MS/s and a laser sweeping rate of 100 kHz. Complex conjugate terms were suppressed in FR-SDM-OCT images with a measured rejection ratio of up to ∼46 dB at ∼1.4 mm depth and ∼30 dB at ∼19.4 mm depth. The feasibility of FR-SDM-OCT was validated by imaging Scotch tapes and human fingernails. Furthermore, we demonstrated the feasibility of FR-SDM-OCT angiography (FR-SDM-OCTA) to perform simultaneous acquisition of human fingernail angiograms from four positions, with a total field-of-view of ∼1.7 mm × ∼7.5 mm. Employing the full-range technique in SDM-OCT can effectively alleviate hardware requirements to achieve the long depth measurement range, which is required by SDM-OCT to separate multiple images at different sample locations. FR-SDM-OCTA creates new opportunities to apply SDM-OCT to obtain wide-field angiography of <i>in vivo</i> tissue samples free of labeling.
radiology, nuclear medicine & medical imaging,biochemical research methods,optics