Optimization of Fiber-Based Spectral Domain Optical Coherence Tomography System

BRAEULER Tobias
DOI: https://doi.org/10.3969/j.issn.1002-2082.2010.03.027
2010-01-01
Journal of Applied Optics
Abstract:The hardware parameters of a spectral domain OCT (SD-OCT) system are analyzed for the optimal OCT imaging. These parameters include pixel number and digitization depth of the line CCD in the spectrometer,and the tolerance of CCD installation. Calibration of the spectrum localization is carried out. The depth resolution is not affected by the CCD linear array pixel number,but with the increasing of CCD linear array pixel number,the maximum probing depth increases linearly. If the digitization depth of CCD is higher than 6bit,it is possible to calculate useful depth cans with good resolutions. When the CCD is moved several millimeters away from the focal plane,the value of point spread function (PSF) will decrease,and the resolution will degrade. When the CCD rotates several degree,full width at half maximum (FWHM) of the PSF will decrease,denoting the increasing of the axial resolution. After calibration using Ne lamp spectrum,the quality of the image will be obviously improved. Some of the results are experimentally verified. For SD-OCT system,the optimal system parameters can be selected for specific application.