Simultaneous Determination of Copper(II) and Cobalt(II) by Ion Chromatography Coupled with Chemiluminescent Detection

Lu Chao,Lin Jin-Ming,Huie Carmen W.,Yamada Masaaki
DOI: https://doi.org/10.2116/analsci.19.557
IF: 1.967
2003-01-01
Analytical Sciences
Abstract:In the absence of any special luminescent reagents, a weakly chemiluminescent emission was observed during the decomposition of hydrogen peroxide, catalyzed by transition-metal ions, such as Cu(II) and Co(II), in basic aqueous solution. The chemiluminescent intensity was significantly enhanced by the addition of ethyldimethylcetylammonium bromide and uranine. The signal-to-noise ratio ( S / N ) was proportional to the concentrations of Cu(II) and Co(II). Based on these phenomena, a flow-injection chemiluminescent method for the simultaneous separation and determination of Cu(II) and Co(II) was developed. The detection limits of the present chemiluminescent method for Cu(II) and Co(II) were 7. 5 and 0. 01 ng/ml, respectively. After ion chromatographic separation of Cu(II) and Co(II) by an IonPac CS5A column with oxalic acid and lithium hydroxide monohydrate as the eluent, the present chemiluminescent system was used as a post-column detector for these two transition metal ions in natural water samples.
What problem does this paper attempt to address?