A Comparison of Corneal Thicknesses Measured Before LASIK by RTVue FD-OCT System and Ultrasonic Pachymetry

赵鹏飞,周跃华,张晶
DOI: https://doi.org/10.3969/j.issn.1004-3594.2010.07.017
2010-01-01
Abstract:Objective To study the dissimilarity and repeatability of the measurement of corneal central thickness(CCT)applied to LASIK preoperative patients between RTVue fourier domain optical coherence tomography(RTVue FD-OCT)and ultrasonic pachymeter(USP).Methods CCT was measured by RTVue FD-OCT and USP separately in 800 myopic preoperative patients (1600 eyes).The eyes were divided into four groups according to the difference in cornea thickness measured by spherical equivalent refraction and USP.These two results of measurement were compared for their difference and correlation.Sixty samples(60 eyes)were selected randomly.The CCT was measured three times using RTVue FD-OCT and USP separately.Results The mean value of RTVue FD-OCT and USP was(528.47±29.84)μm and(537.89±30.56)μm,respectively;the difference was statistically significant (P0.01).RTVue and USP were in linear correlation(r=0.921,P0.01).Their relevant formula was:the measurement value of OCT=2.2492×USP measurement value 0.8678.The mean standard deviations of repeated CCT measurement value of OCT and USP were 3.87 and 6.38 respectively.The discrepancy of standard deviation between RTVue and USP was of statistical significance(P 0.01).Conclusions RTVue FD-OCT is as efficient in measuring the corneal thickness as ultrasound.Thus,it can be used as a tool for pachymetry and a method for meta-analysis and observation before and after LASIK.