[Determination of Plasma Escitalopram with Liquid Chromatography-Mass Spectrometry].

Heng Zheng,Ling Yang,Wen-Yan Tang
DOI: https://doi.org/10.3321/j.issn:1673-4254.2008.11.053
2008-01-01
Abstract:OBJECTIVE:To establish an liquid chromatography-mass spectrometry (HPLC-MS/MS)-based method for determining escitalopram in human plasma.METHODS:A liquid-liquid ether extraction approach was adopted using dextromethorphan as the internal standard. The Agilent ZORBAX SB-C18 (3.5 microm,2.1 x 150 mm) was used as the analytical column with acetonitrile-NH4AC buffer (70:30, V/V) as the mobile phase at the flow rate of 0.3 ml/min. The sample was ionized by electrospray ionization source in the triple quadruple tandem mass spectrometer, and the plasma escitalopram determined with a multiple reaction monitoring mode of m/z325.4-->109.4.RESULTS:The linear range was 0.0866-64.13 microg/L (r=0.9965) for escitalopram in human plasma, with the absolute recovery between 64.98% and 78.72% and the within-day and between-day deviations less than 10%.CONCLUSION:The method is sensitive, accurate, rapid, specific and well applicable for clinical pharmacokinetics study of escitalopram.
What problem does this paper attempt to address?