Determination of Sertraline Concentration in Healthy Human Plasma by HPLC

WANG Ping,CHEN Xing-yan,QIU Xiang-jun,HU Guo-xin,CHEN Bing-bing,XU Peng
DOI: https://doi.org/10.3969/j.issn.1001-6821.2006.01.017
2006-01-01
Abstract:Objective To establish a high performance liquid chromatography method for the determination of sertraline in healthy human plasma.Methods Agilent 1100 series HPLC system was applied with the column Zorbax Extend-C_ 18 (4.6 mm×250 mm,5 μm).The mobile phase was composed of methanol (0.15 mol·L -1 ) and NaH_2PO_4(pH 3.0) with ratio of 40∶60,and the flow rate was 0.80 mL· min -1 . The UV detection wavelength was 230 nm. Results The calibration curve was liner in the range from 1.0 to 150.0 μg·L -1 (γ= 0.9999) ,with determination limit of 1 μg·L -1 and the average recovery of (100.54 ± 4.01 ) %. The intra-day RSD and inter-day RSD were all less than 5%. Conclusion The method was simple ,rapid , accurate and can be used to determine the sertraline concentration in plasma.