Amperometric glucose biosensor based on carbon nanotubes and platinum nanoparticle modified gold electrode

Duan Da-xue,Chu Xia
DOI: https://doi.org/10.3969/j.issn.1008-2298.2006.02.006
2006-01-01
Chemical Sensors
Abstract:With glutaraldehyde acting as cross linker, glucose oxidase has been immobilized on polylysine modified carbon nanotubes, then the glucose biosensor has been developed by laying polymer Nafion and carbon nanotubes, which has been cross linked with GOx, on platinum nanoparticles modified gold electrode. The effect of Pt deposition time, the amout of CNT in Nafion, pH of the buffer solution and working potential on the response of the electrode has been investigated. The experimental results showed that the designed biosensor has good stability and rapid response. The catalytic current is linearly related to the glucose concentration in the range of 0.1μmol/L ~6 mmol/L, and the response time is less than 5 s.
What problem does this paper attempt to address?