Experimental Study of the Effect of Heart Rate on the Image Quality of Coronary Artery with 64-detector Row Helical CT Using a Pulsating Cardiac Phantom

LIU Bin,ZHAO Hong,WU Xing-wang
DOI: https://doi.org/10.3969/j.issn.1000-0313.2007.05.007
2007-01-01
Abstract:Objective:Using a pulsating cardiac phantom to study the effect of heart rate on the image quality of coronary artery with 64-detector row helical CT.Methods:An adjustable pulsating cardiac phantom (GE) containing predetermined simulated coronary arteries was scanned using 64-detector row helical CT with cardiac pulsating rates as 40,45,50,55,60,65,70,75,80,85,90,95,100,105,110 and 115 beats per minute (bpm).The rotation speed of 0.35s was used.The raw data were reconstructed at an optimal window of the R-R interval using both one sector and multi-sector and volume rendering (VR) and multi-planar reformation (MPR) as the reconstruction algorithm.The image quality score (IQS) was evaluated by two radiologists respectively.Results:There was negative correlation between heart rate and IQS (P<0.01).The image quality of multi-sector reconstruction algorithm was superior than that of one-sector.Conclusion:Using a pulsating cardiac phantom,with 64-detector row CT,the effect of heart rate on the image quality of coronary artery could be evaluated,which plays an important role in clinical research as well as clinical application.