Effects of heart rate and heart rate variability on image quality and radiation exposure of 64-slice coronary CT angiography

Jinghong He,Jianding Li,Ying Qiao,Ruiping Zhang
DOI: https://doi.org/10.16106/j.cnki.cn14-1281/r.2016.01.001
2016-01-01
Abstract:Objective To assess the effect of heart rate and heart rate variability on image quality and patient dose with 64-slice sprial CT coronary angiography. Methods Ninty patients underwent 64-slice coronary CT angiog-raphy.Image quality was performed usding four-score method.The analysis was performed to evaluate the relationship of heart rate and heart rate variability on the image quality and radiation dose. Results Three hundreds and sixty coronary arteries were analyzed in 90 patients, 97.5%of the coronary segments was sufficient for diagnosing. The aver-age heart rate was (67 ±10) beat per minute,with a variability of (7.2 ±9.3)%.The mean effective dose was (20.2 ±2.3) mSv.Significant negative correlations were observed both between heart rate and image quality and between heart rate variability and image quality. The results showed that heart rate and heart rate variability had no significant effect on radiation dose. Conclusion The image quality was significantly different in patients with different heart rate and heart rate variability.Reducing average heart rate and heart rate variability in patients is beneficial in improving image quality with 64-slice sprial CT coronary angiography(r1=-0.651, P<0.001; r2=0.662, P<0.001),but have no help to re-duce radiation dose.