Analysis of the reliabitity of the abnormal alarming parameter of ADVIA120 hematology analyzer

AN Bangquan,LING Xiaowu,ZHANG Yan,ZHOU Xianghong,YE Aizhu,ZHANG Jian,FEI Rong
DOI: https://doi.org/10.3969/j.issn.1673-8640.2006.05.019
IF: 1.865
2006-01-01
Laboratory Medicine
Abstract:Objective To study the reliability of the function of ADVIA120 automatic hematology analyzer on abnormal shape of cells. Methods 160 samples with abnormal parameter hinted by the analyzer and 112 samples without hinted abnormal parameter were tested comparative with microscopic examination of blood smears after Wright's staining. Results True positive rate: immature granulocyte(IG),hemoglobin concentration variation(HC-VAR),microcyte(MICRO),left shift(LS),macrocyte(MACRO),hypochromic erythrocyte(HYPO),large platelet(LPLT),blasts,anisocytosis(ANISO) 86.7%-100.0%,hyperchromic erythrocyte(HYPER) 69.2%,nucleated red blood cell(NRBC),atypical lymphocyte(ATYP),platelet clumps(PLTCLM)(13.3%)-32.3%.All the parameters above had a true negative rate of 82.8%-100.0%(mean of 95.0%). Conclusions The analyzer has a high true negative rate towards alarming parameter and has a strong screening power.The abnormal alarming message on NRBC,ATYP,PLTCLM produced by this analyzer is not accurate ideally,but the abnormal alarming message of other parameters is reliable to a great extent,therefore the samples with the outcome hinted abnormal should be compared with microscopic observation to find the diagnosis.