Establishment of the Magnetic Separation Enzyme Immunofluorescence Method for Detecting Insulin

Jing Xie,Zhen-shi Liu,Shan Hu,Ru-xiu Cai,Hai Yang,Xiang-liang Yang
DOI: https://doi.org/10.3321/j.issn:1007-8738.2008.04.014
2008-01-01
Abstract:AIM:To establish a convenient and sensitive magnetic separation enzyme immunofluorescence (MEIF) method for detecting human insulin.METHODS:Two monoclonal antibodies (mAbs) were conjugated with FITC and alkaline phosphatase (AP) respectively, which were incorporated magnetic solid phase separation. Magnetic beads were coupled with sheep anti-FITC antibody as solid phase and 4-Methylumbelliferyl-phosphoric acid (4-MUP) was used as substrate to set up MElF for detecting insulin.RESULTS:The sensitivity of this assay was 2.0 microIu/mL, the linear range was from 0 microIu/mL to 188.52 microIu/mL, and the intra-assay variation and inter-assay variation were 4.3%-5.2% and 2.6%-9.5%, respectively. The recovery rate of dilution was 92.6%-117% and the recovery rate of accession was 106%-121%.The result of the assay correlated well with that of magnetic enzyme chemiluminescence immunoassay system.CONCLUSION:The MEIF for detecting insulin is low at cost, sensitive, specific and stable, which can be widely used in clinical immune detection.