Determination of Carcinoembryonic Antigen Based on Magnetic Microsphere Assited Fluoroimmunoassay

Rong LI,Feng FENG,Ze-zhong CHEN,Yun-feng BAI,Fang-fang GUO,Fang-ying WU,Gao ZHOU
DOI: https://doi.org/10.3969/j.issn.1674-0874.2015.04.011
2015-01-01
Abstract:A novel immunoassay for the determination of CEA was established by combining a fluoroimmunoassay and immuno?magnetic separation. Carboxyl groups on the magnetic beads were first activated with a mixture of EDC and NHS to give reactive succin?imide esters. The mAbCEA-C3 was then passed over the surface and esters react spontaneously with amino groups to link mAbCEA-C3 covalently to magnetic beads to form Immonumagnetic beads(IBMs). In the mixture samples, CEA was captured by mAbCEA-C 3 with high specificity and affinity. The conditions, such as washing times and immune reaction time were optimized. The final choice of immune time is 45min, three times of washing. The results indicated that the fluorescence intensity increased with the increment of CEA concentration in the optimal condition. The LOD is 9.67 ng/mL. Qualitative and semiquantitative detection of CEA was conducted based on the combination of IBMs with fluoroimmunoassay. The method is simple and sensitive and could be a possible application for the detection of CEA in clinical diagnostics.
What problem does this paper attempt to address?