Micro-plate magnetic chemiluminescence immunoassay and its applications in carcinoembryonic antigen analysis

ZhiYong Li,QianYun Zhang,LiXia Zhao,ZhenJia Li,GuoMao Hu,JinMing Lin,Shan Wang
DOI: https://doi.org/10.1007/s11426-009-0168-y
2009-01-01
Science China Chemistry
Abstract:A micro-plate magnetic chemiluminescence immunoassay was developed for rapid and high throughput detection of carcinoembryonic antigens (CEA) in human sera. This method was based on a sandwich immunoreaction of fluorescein isothiocyanate (FITC)-labeled anti-CEA antibodies, CEA antigens, and horseradish peroxidase (HRP)-conjugated anti-CEA antibodies in micro-plate. The immunomagnetic particles coated with anti-FITC antibodies were used as the solid phase for the immunoassay. The separation procedure was carried out by a magnetic plate adaptor and the luminol-hydrogen peroxide (H 2 O 2 )-HRP system was employed for the chemiluminescence detection. The proposed method combined the advantages of the micro-plate reactor and magnetic particle separation technology with the linear range of 5–250 ng·mL −1 . The detection limit of CEA was 0.61 ng·mL −1 . The coefficient of the variation was less than 7% and 13% for intra-assay and inter-assay precision, respectively. Compared with the commercial micro-plate chemiluminescent kit, the proposed method showed a good correlation.