Establishment and evaluation of a flow cytometric immune microsphere assay for quantitative detecting human serum insulin

WANG Zhanke,HE Dan,ZHU Zhongzhen,HU Xiaolu,PAN Xiaoqing,CHANG Jin
DOI: https://doi.org/10.13431/j.cnki.immunol.j.20120078
2012-01-01
Abstract:We aim to establish a novel flow cytometric immune microsphere assay(FCIMA) for quantitative detecting human serum insulin by flow cytometric immune microsphere assay(FCIMA) and study its repeatability,linearity,and lowest detectable limit(LDL).Microsphere surface were labeled by rabbit against human insulin monoclonal antibody to prepare immune diagnosis microspheres,which could be combined with human insulin of serum,and appear fluorescent light by combining rat against human insulin monoclonal antibody labeled with FITC.The mean fluorescent intensity(MFI) of the surface of immune diagnosis microspheres were detected by flow cytometer(FC) and recorded.We found that the MFI of immune diagnosis microspheres were in direct proportion to the concentrations of human serum insulin.Moreover,both of the coefficient of variance(CV) of repeatability and the LDL of FCIMA were lower than those of ELISA,and the range of linearity of FCIMA was wider than that of ELISA.These results indicated that FCIMA can be used for quantitative determination of human serum insulin,which is better than ELISA in the repeatability,linearity range,and detection sensitivity.
What problem does this paper attempt to address?