Experimental Investigation on Range Extension of GM Tube by Time-to-Count Method

GONG Jun-jun,SHAO Bei-bei,XIA Wen-ming,CHEN Jun-jun
DOI: https://doi.org/10.7538/yzk.2013.47.04.0674
2013-01-01
Abstract:GM tube is a kind of widely used nuclear radiation detector, whose linear range is rather narrow.Major drawbacks were analyzed by adopting dead time models and function fitting.The principle of Time-to-Count method was introduced.The design scheme of an experimental apparatus based on Time-to-Count method was presented.The problems such as pulse-controlled high voltage driving and time interval measuring were discussed.Experimental apparatus was tested in National Defense Science Technology Ionizing Radiation First-order Metrology Station and test data show that the range is extended to 5orders of magnitude and non-linearity is less than 10%in the whole range.
What problem does this paper attempt to address?