P257 UNIQUE UNEXPECTED COURSE OF A S–ICD LEAD: DIAGNOSTIC CHALLENGES AND SAFE CLINICAL AND INTERVENTIONAL MANAGEMENT
C Cataldi,A Persi,F Percoco,M Bacà,F Perna,G Pelargonio,C Micheli,S Canitano,A Kol
DOI: https://doi.org/10.1093/eurheartjsupp/suad111.332
2023-05-01
European Heart Journal Supplements
Abstract:Abstract In this report we present a case of subcutaneous cardioverter defibrillator (S–ICD) implantation complicated by pneumothorax (PNX), in the absence of alteration of electrical parameters nor lung parenchyma involvement. After implantation in secondary prevention (1995) of an abdominal transvenous defibrillator (TV–ICD) via the left subclavian vein, a replacement (2009) with a new contralateral TV–ICD system, and the withdrawal of the proximal edge of the right ventricular (RV) lead in the abdominal region, a 78–years–old man presented in our center for pocket erosion. The trans–esophageal echocardiography was suspect for vegetations, while PET/CT and blood cultures were negative. Transvenous extraction of the right–sided ICD system was performed (the original RV lead was left in place due to his complex course), then the patient underwent implantation of a S–ICD with the intermuscular 2–incision technique. In the end of this well tolerated procedure the electrical parameters were optimal, the defibrillation test was effective and the position of the S–ICD was judged correct on the chest X–ray. Few hours later, the patient complained for local pain and progressive respiratory failure. A chest CT showed a PNX at the lateral insertion of the lead, which reached the pleural space, along the left scissure (Fig.1). During lead repositioning, programmed with stand–by thoracic surgery, the lead was easily disconnected from the generator through the lateral incision, pulled–back through the subxiphoid one and tunneled in the correct plane without complications (Fig.2). Serial chest CT confirmed PNX resolution. The patient was discharged in 8 days. At the 1–month follow–up visit electrical parameters were stable, in absence of infective complications. In literature there are few cases of PNX, so that it’s challenging to define their risk factors. We assume that, despite a correct implantation technique, a barrel chest configuration and the aging of the fascia favored the wrong tunnelling. In conclusion, PNX is a rare and unexpected complication of S–ICD implantation and can go unnoticed when of limited dimensions, in the absence of clinical and radiological signs nor alteration of electrical parameters. Fluoroscopic and CT imaging is recommended for diagnosis. Thoracic surgery is useful during repositioning of the lead, which can also be performed by simple traction. CT imaging is of utmost importance for planning such a procedure and predicting its complexity.
cardiac & cardiovascular systems