Clinical analysis of the common complications in permanent cardiac pacemaker
MA Yi-tong,LIU Zhi-qiang,TANG Bao-peng,SHI Xue-kun,KUANG Zheng-yu
2008-01-01
Abstract:Objective To explore the causes of common complications in permanent pacemaker and its preventive and treatment measures.Methods From January 2005 to April 2008 complications of 456 patients with pacemaker(male 275,female 181,aged 28-91 years,mean age 60.23 years)were retrospectively analyzed.Among them there were 287 cases of sick sinus syndrome,97 Ⅲ° AVB,28 high degree AVB,19Ⅱ° AVB,12 dilated cardiomyopathy,and 13 with other diseases.The types of installed pacemakers were single-chamber pacemaker,dual-chamber pacemaker,three-chamber pacemaker and inplantable cardiac defibrilator(ICD).Results The patients with pacemaker complications were 27 cases after permanent cardiac pacemaker implantation,4 of incorrect puncture in clavicle artery,1 of pneumothorax,1 of coronary vein dissection,1 of injecting bubbles into coronary vein,1 of drain strip split,2 of electrode reverse linking,2 of perception disturbance,1 of battery exhausting,2 of pacemaker syndrome,1 of pacemaker interventional tachycardia,4 of electrode dislocation,6 of pocket hematoma and 1 of pocket infection.Conclusion Together of the following measures eg.throughly understanding of the common complications,increased differential ability,paying attention to preoperative prevention,standardized operation procedure,closed following up as well as strengthen the education of pacemaker knowledge and early discovery and treatment of the complications would significantly decrease the complication occurrence rate of pacemaker implantation.