Super-resolution deep-learning reconstruction for cardiac CT: impact of radiation dose and focal spot size on task-based image quality

Takafumi Emoto,Yasunori Nagayama,Sentaro Takada,Daisuke Sakabe,Shinsuke Shigematsu,Makoto Goto,Kengo Nakato,Ryuya Yoshida,Ryota Harai,Masafumi Kidoh,Seitaro Oda,Takeshi Nakaura,Toshinori Hirai
DOI: https://doi.org/10.1007/s13246-024-01423-y
2024-06-17
Abstract:This study aimed to evaluate the impact of radiation dose and focal spot size on the image quality of super-resolution deep-learning reconstruction (SR-DLR) in comparison with iterative reconstruction (IR) and normal-resolution DLR (NR-DLR) algorithms for cardiac CT. Catphan-700 phantom was scanned on a 320-row scanner at six radiation doses (small and large focal spots at 1.4-4.3 and 5.8-8.8 mGy, respectively). Images were reconstructed using hybrid-IR, model-based-IR, NR-DLR, and SR-DLR algorithms. Noise properties were evaluated through plotting noise power spectrum (NPS). Spatial resolution was quantified with task-based transfer function (TTF); Polystyrene, Delrin, and Bone-50% inserts were used for low-, intermediate, and high-contrast spatial resolution. The detectability index (d') was calculated. Image noise, noise texture, edge sharpness of low- and intermediate-contrast objects, delineation of fine high-contrast objects, and overall quality of four reconstructions were visually ranked. Results indicated that among four reconstructions, SR-DLR yielded the lowest noise magnitude and NPS peak, as well as the highest average NPS frequency, TTF50%, d' values, and visual rank at each radiation dose. For all reconstructions, the intermediate- to high-contrast spatial resolution was maximized at 4.3 mGy, while the lowest noise magnitude and highest d' were attained at 8.8 mGy. SR-DLR at 4.3 mGy exhibited superior noise performance, intermediate- to high-contrast spatial resolution, d' values, and visual rank compared to the other reconstructions at 8.8 mGy. Therefore, SR-DLR may yield superior diagnostic image quality and facilitate radiation dose reduction compared to the other reconstructions, particularly when combined with small focal spot scanning.
What problem does this paper attempt to address?