X-ray counting efficiencies for plutonium in lungs, derived from studies with inhaled palladium-103.

D. Newton,B. T. Taylor,A. L. Anderson
DOI: https://doi.org/10.1097/00004032-197806000-00006
1978-06-01
Health Physics
Abstract:Abstract A semi-empirical model of the human thorax has been developed to predict the influence of photon energy on the efficiency of external counting of low-energy X-ray emitters in lungs. The calculations allow for differential attenuation of X-rays according to photon energy both within the lungs and in the chest wall. The model is specifically employed to derive calibration data for estimating accidentally inhaled 239Pu from known X-ray detection efficiencies for 103Pd inhaled by volunteers; however, it is also potentially useful in the assessment of other low-energy photon emitters. The influence of anatomical and other factors on the relative detection efficiency, R, for 103Pd and 239Pu is investigated. The dominant factor in the model is the thickness of the chest wall in the regions viewed by the detectors, but the unknown pattern of distribution of activity within the lungs would be the largest single source of uncertainty in R. In general, a uniform distribution is assumed, leading to an error which would probably not exceed 20% unless the actual distributions were grossly uneven. Experimental evidence of the validity of the model in this application is provided, and calibration factors are derived for assessing 239Pu in lungs with phoswich detectors of 12 cm diameter.