Balanced-Detection Interferometric Cavity-Assisted Photothermal Spectroscopy via Collimating Fiber-Array Integration
Giulia Malvicini,Johannes P. Waclawek,Davide Pinto,Harald Moser,Simone Iadanza,Kamil Gradkowski,Liam O’Faolain,Bernhard Lendl
DOI: https://doi.org/10.1016/j.snb.2024.135766
IF: 9.221
2024-04-07
Sensors and Actuators B Chemical
Abstract:This work reports on the implementation of a fiber-array with attached microlenses in the balanced-detection Interferometric Cavity-Assisted Photothermal Spectroscopy (ICAPS) scheme, enabling a highly compact probe laser beam path. The sample probe and the reference beam were coupled into a 1-mm spaced Fabry-Perot interferometer with a parallel beam separation of only 1.25 mm. A quantum cascade laser was used to induce photo-thermal waves, which were detected via a sample probe beam. A reference probe beam was used to realize balanced detection, allowing for significant noise reduction. The sensor's metrological figures of merit were evaluated by detection of nitric oxide. For the targeted absorption band centered at 1900.08 cm -1 a 1 σ minimum detection limit of 64 ppbv was achieved with a 1-second integration time. This performance corresponds to a normalized noise equivalent absorption of 5.7 ×10 -8 cm -1 W Hz -1/2 .
chemistry, analytical,electrochemistry,instruments & instrumentation