Dosimetry of microbeam radiotherapy by flexible hydrogenated amorphous silicon detectors
Matthew James Large,Keida Kanxheri,Jessie Posar,Saba Aziz,Aishah Bashiri,Lucio Calcagnile,Daniela Calvo,Domenico Caputo,Anna Paola Caricato,Roberto Catalano,Roberto Cirio,Giuseppe Antonio Pablo Cirrone,Tommaso Croci,Giacomo Cuttone,Gianpiero De Cesare,Paolo De Remigis,Sylvain Dunand,Michele Fabi,Luca Frontini,Catia Grimani,Mariacristina Guarrera,Maria Ionica,Francesca Lenta,Valentino Liberali,Nicola Lovecchio,Maurizio Martino,Giuseppe Maruccio,Giovanni Mazza,Mauro Menichelli,Anna Grazia Monteduro,Arianna Morozzi,Francesco Moscatelli,Augusto Nascetti,Stefania Pallotta,Daniele Passeri,Maddalena Pedio,Giada Petringa,Francesca Peverini,Pisana Placidi,Gianluca Quarta,Silvia Rizzato,Federico Sabbatini,Leonello Servoli,Alberto Stabile,Jonathan Emanuel Thomet,Luca Tosti,Mattia Villani,Richard James Wheadon,Nicolas Wyrsch,Nicola Zema,Marco Petasecca,Cinzia Talamonti
DOI: https://doi.org/10.1088/1361-6560/ad64b5
2024-07-26
Abstract:Objective.Detectors that can provide accurate dosimetry for microbeam radiation therapy (MRT) must possess intrinsic radiation hardness, a high dynamic range, and a micron-scale spatial resolution. In this work we characterize hydrogenated amorphous silicon detectors for MRT dosimetry, presenting a novel combination of flexible, ultra-thin and radiation-hard features.Approach.Two detectors are explored: an n-type/intrinsic/p-type planar diode (NIP) and an NIP with an additional charge selective layer (NIP + CSC).Results.The sensitivity of the NIP + CSC detector was greater than the NIP detector for all measurement conditions. At 1 V and 0 kGy under the 3T Cu-Cu synchrotron broadbeam, the NIP + CSC detector sensitivity of (7.76 ± 0.01) pC cGy-1outperformed the NIP detector sensitivity of (3.55 ± 0.23) pC cGy-1by 219%. The energy dependence of both detectors matches closely to the attenuation coefficient ratio of silicon against water. Radiation damage measurements of both detectors out to 40 kGy revealed a higher radiation tolerance in the NIP detector compared to the NIP + CSC (17.2% and 33.5% degradations, respectively). Percentage depth dose profiles matched the PTW microDiamond detector's performance to within ±6% for all beam filtrations except in 3T Al-Al due to energy dependence. The 3T Cu-Cu microbeam field profile was reconstructed and returned microbeam width and peak-to-peak values of (51 ± 1)μm and (405 ± 5)μm, respectively. The peak-to-valley dose ratio was measured as a function of depth and agrees within error to the values obtained with the PTW microDiamond. X-ray beam induced charge mapping of the detector revealed minimal dose perturbations from extra-cameral materials.Significance.The detectors are comparable to commercially available dosimeters for quality assurance in MRT. With added benefits of being micron-sized and possessing a flexible water-equivalent substrate, these detectors are attractive candidates for quality assurance,in-vivodosimetry and in-line beam monitoring for MRT and FLASH therapy.