3D Prompt Gamma Imaging in a Dual-head Multi-slit System for Proton Beam Range Monitoring

H. Zhang,B. Zhao,Z. Wang,P. Fan,S. Wang,Z. Wu,T. Ma,Y. Liu
DOI: https://doi.org/10.1109/nssmicrtsd49126.2023.10338062
2023-01-01
Abstract:Proton range online verification is of great importance in proton therapy due to range uncertainty. Prompt gamma imaging is a promising method for online range monitoring.In previous work, a dual-head system aiming at 3D prompt gamma imaging was already assembled and calibrated. Spatial resolution of the system is better than 2 mm and sensitivity at the center of the field of view (FOV) is 0.62%, according to results in 60 Co point source experiment. In this work, the first system test under proton therapy equipment is reported. 100 MeV, 0.5 nA proton beam is used in the experiment. Prompt gamma generated by a polymethylmethacrylate (PMMA) phantom is collected by the system. A maximum-likelihood expectation-maximization (MLEM) algorithm is used for 3D prompt gamma image reconstruction.Range shift of 1 mm is able to be distinguished by calculating the most distal edge position (MDEP) of prompt gamma profile. MDEP accuracy evaluated through repeated experiments is 0.2 mm. Positioning accuracy of the prompt gamma distribution peak position in proton beam cross section is 0.2 mm. Range deviation accuracy calculated by prompt gamma image of 1×10 8 protons is less than 1 mm.The dual-head imaging system provides excellent result when the Bragg peak of proton beam is located near the center of the FOV. Proton range resolving ability of the system in proton pencil beam scanning lattice will be evaluated in the future.