Design and calibration of a cryogenic blackbody calibrator at centimeter wavelengths

A. Kogut,E. Wollack,D. J. Fixsen,M. Limon,P. Mirel,S. Levin,M. Seiffert,P. M. Lubin
DOI: https://doi.org/10.1063/1.1821622
IF: 1.6
2004-12-01
Review of Scientific Instruments
Abstract:We describe the design and calibration of an external cryogenic blackbody calibrator used for the first two flights of the Absolute Radiometer for Cosmology, Astrophysics, and Diffuse Emission (ARCADE) instrument. The calibrator consists of a microwave absorber weakly coupled to a superfluid liquid helium bath. Half-wave corrugations viewed 30° off axis reduce the return loss below −35 dB within a compact footprint. Ruthenium oxide resistive thermometers embedded within the absorber monitor the temperature across the face of the calibrator. The thermal calibration transfers the calibration of a reference thermometer to the flight thermometers using the flight thermometer readout system. The calibrator thermometry is stable in time over four years, with statistical uncertainty in the temperature calibration of order 2 mK near 2.7 K, limited primarily by thermal fluctuations in the liquid helium bath. Observations of the superfluid transition demonstrate that the absolute temperature scale is accurate within 0.3 mK.
instruments & instrumentation,physics, applied