A cryogenic infrared calibration target

E J Wollack,R E Kinzer,S A Rinehart,E. J. Wollack,R. E. Kinzer,S. A. Rinehart
DOI: https://doi.org/10.1063/1.4871108
IF: 1.6
2014-04-01
Review of Scientific Instruments
Abstract:A compact cryogenic calibration target is presented that has a peak diffuse reflectance, R ⩽ 0.003, from 800 to 4800 cm(-1) (12 - 2 μm). Upon expanding the spectral range under consideration to 400-10,000 cm(-1) (25 - 1 μm) the observed performance gracefully degrades to R ⩽ 0.02 at the band edges. In the implementation described, a high-thermal-conductivity metallic substrate is textured with a pyramidal tiling and subsequently coated with a thin lossy dielectric coating that enables high absorption and thermal uniformity across the target. The resulting target assembly is lightweight, has a low-geometric profile, and has survived repeated thermal cycling from room temperature to ∼4 K. Basic design considerations, governing equations, and test data for realizing the structure described are provided. The optical properties of selected absorptive materials-Acktar Fractal Black, Aeroglaze Z306, and Stycast 2850 FT epoxy loaded with stainless steel powder-are characterized and presented.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?