14 C AMS measurement of CO 2 samples with the high throughput HVE gas interface system

G Scognamiglio,M Klein,A Stolz,R Walet,D Mous
DOI: https://doi.org/10.1017/rdc.2024.60
IF: 6.324
2024-09-19
Radiocarbon
Abstract:The newly designed HVE gas interface enables the AMS measurement of carbon samples in CO 2 form. The CO 2 , e.g. resulting from the sample combustion in an elemental analyzer, is adsorbed in a zeolite trap and subsequently transferred to a motor-driven syringe. Once diluted with He, the gas mixture is transferred into the ion source of the AMS system. A carbon ion beam is formed in the ion source and mass-analyzed by the AMS system, resulting in 13 C/ 12 C and 14 C/ 12 C isotopic ratios. The HVE gas interface features two traps and two syringes to maximize the sample throughput, which results in more than 10 samples per hour. The first performance results of CO 2 gas sample AMS measurements that were performed with the HVE gas interface in combination with the HVE 210 kV AMS system are presented in this paper. The measurements show that the gas interface contribution to the 14 C/ 12 C background is in the 10 –15 level and to the precision is at or below 1%.
geochemistry & geophysics
What problem does this paper attempt to address?