Pull-out resistance of connected K-wires for osteosynthesis: development of a numerical model

Samuel BertheNadia BahlouliMassamaesso BilassePhilippe Liverneauxa ICube CNRS UMR7357,Strasbourg University,Strasbourg,Franceb ICAM,Schiltigheim,Francec Department of Hand Surgery,Strasbourg University Hospitals,FMTS,Strasbourg,France
DOI: https://doi.org/10.1080/10255842.2024.2310729
IF: 1.6
2024-01-31
Computer Methods in Biomechanics & Biomedical Engineering
Abstract:A predictive finite element model was developed to investigate the best configuration of a fixation pins system consisting of two K-wires inserted in a synthetic model (Sawbones ® ) at different angles and secured to a connecting rod. Two key parameters were considered to determine the best configuration delivering the higher pull-out strength and lower pull-out length: the diameter and insertion angle. Results show that as the diameter and insertion angle increased, the pull-out force increased, while the pull-out length decreased. Results are successfully compared with available experimental data in literature. This model can be used as an alternative to experimental study.
engineering, biomedical,computer science, interdisciplinary applications
What problem does this paper attempt to address?