A magnetically shielded room with ultra low residual field and gradient
I Altarev,E Babcock,D Beck,M Burghoff,S Chesnevskaya,T Chupp,S Degenkolb,I Fan,P Fierlinger,A Frei,E Gutsmiedl,S Knappe-Grüneberg,F Kuchler,T Lauer,P Link,T Lins,M Marino,J McAndrew,B Niessen,S Paul,G Petzoldt,U Schläpfer,A Schnabel,S Sharma,J Singh,R Stoepler,S Stuiber,M Sturm,B Taubenheim,L Trahms,J Voigt,T Zechlau,I. Altarev,E. Babcock,D. Beck,M. Burghoff,S. Chesnevskaya,T. Chupp,S. Degenkolb,I. Fan,P. Fierlinger,A. Frei,E. Gutsmiedl,S. Knappe-Grüneberg,F. Kuchler,T. Lauer,P. Link,T. Lins,M. Marino,J. McAndrew,B. Niessen,S. Paul,G. Petzoldt,U. Schläpfer,A. Schnabel,S. Sharma,J. Singh,R. Stoepler,S. Stuiber,M. Sturm,B. Taubenheim,L. Trahms,J. Voigt,T. Zechlau
DOI: https://doi.org/10.1063/1.4886146
IF: 1.6
2014-07-01
Review of Scientific Instruments
Abstract:A versatile and portable magnetically shielded room with a field of (700 ± 200) pT within a central volume of 1 m × 1 m × 1 m and a field gradient less than 300 pT/m, achieved without any external field stabilization or compensation, is described. This performance represents more than a hundredfold improvement of the state of the art for a two-layer magnetic shield and provides an environment suitable for a next generation of precision experiments in fundamental physics at low energies; in particular, searches for electric dipole moments of fundamental systems and tests of Lorentz-invariance based on spin-precession experiments. Studies of the residual fields and their sources enable improved design of future ultra-low gradient environments and experimental apparatus. This has implications for developments of magnetometry beyond the femto-Tesla scale in, for example, biomagnetism, geosciences, and security applications and in general low-field nuclear magnetic resonance (NMR) measurements.
instruments & instrumentation,physics, applied