A Method for Stabilizing Magnetic Field

WANG Siguang,Bouwhuis M,Holler Y,Lenisa p,Reggiani D,Marukyan H,Tait P,YE Zhenyu,YE Yunxiu
DOI: https://doi.org/10.3321/j.issn:0253-3219.2005.04.011
2005-01-01
Nuclear Techniques
Abstract:The magnetic flux density of the HERMES transverse polarized gas target changing with time and an automatic compensating system for stabilizing the strong magnetic field are described in this paper. The instability of magnetic field was found to be related to temperature change of the ferromagnetics. With the automatic system, the field in the monitoring point around 338.6 mT can be stabilized with the accuracy of ±0.004 mT(Stat) or ±0.045 mT (Syst) instead of ±0.45 mT without the system.
What problem does this paper attempt to address?