A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space

Luis Manuel Montaño-Zetina,Omar Villalobos-Mora
DOI: https://doi.org/10.3791/52272
2016-02-01
Abstract:A simple Positron Emission Tomography (PET) prototype has been constructed to fully characterize its basic working principles. The PET prototype was created by coupling plastic scintillator crystals to photomultipliers or PMT's which are placed at opposing positions to detect two gamma rays emitted from a radioactive source, of which is placed in the geometric center of the PET set-up. The prototype consists of four detectors placed geometrically in a 20 cm diameter circle, and a radioactive source in the center. By moving the radioactive source centimeters from the center the system one is able to detect the displacement by measuring the time of flight difference between any two PMT's and, with this information, the system can calculate the virtual position in a graphical interface. In this way, the prototype reproduces the main principles of a PET system. It is capable to determine the real position of the source with intervals of 4 cm in 2 lines of detection taking less than 2 min.