An Exploration of Parallel Imaging System for Very-low Field (50mT) MRI Scanner

Lei Yang,Wei He,Sheng Shen,Yucheng He,Jiamin Wu,Zheng Xu
2024-11-12
Abstract:Reducing the scanning time of very-low field (VLF) magnetic resonance imaging (MRI) scanners, commonly employed for stroke diagnosis, can enhance patient comfort and operational efficiency. The conventional parallel imaging (PI) technique for high-field MRI should be tailored to apply here, considering the differences in the direction of the main magnetic field and the presence of noise. A VLF-specific PI algorithm and phased-array coil are proposed, marking the first application of PI in VLF MRI. Reconstruction quality is enhanced by denoising undersampled k-space data using a linear-prediction based Kalman filter. Subsequently, the denoised k-space data are nonlinearly mapped from the original space onto a high-dimensional feature space, utilizing a polynomial feature mapping defined nonlinear frame. Frame parameters are calculated using auto-calibration signals (ACS) from the center k-space, and missing phase-encoding lines in the original space are estimated using acquired lines in the feature space. An 8-channel phased-array coil, designed for a vertical main magnetic field, is decoupled using geometric overlap and a low input impedance (LII) preamplifier. Healthy volunteer head imaging experiments using the proposed PI technique exhibit the lowest mean-squared-error (MSE) value and the highest peak-signal-to-noise (PSNR) and structural similarity index (SSIM) values compared to two widely used PI methods. The proposed PI technique enables the VLF MRI scanner to achieve similar image quality and a 72.5% improvement in signal-to-noise ratio (SNR) compared to fully sampled images while requiring less than 50% of the scan time. We present a PI technique tailored for VLF MRI scanner for the first time, along with potential research direction to achieve greater reduction factor.
Medical Physics,Instrumentation and Detectors
What problem does this paper attempt to address?
The problem that this paper attempts to solve is: in very - low - field (50mT) magnetic resonance imaging (VLF MRI) scanners, due to the significant reduction in the main magnetic field strength, the magnetic resonance signal is weakened, and thus multiple scans are required to obtain satisfactory image quality. This not only prolongs the imaging time but also affects the patient's comfort and the operating efficiency of the scanner. To solve this problem, the paper proposes a new parallel imaging technique, including a specially - designed parallel imaging algorithm and an 8 - channel phased - array coil, to achieve accelerated imaging and improve image quality. ### Specific problem description: 1. **Weak magnetic resonance signal**: - In very - low - field MRI, due to the low main magnetic field strength, the magnetic resonance signal is significantly weakened. - In order to obtain high - quality images, multiple scans are usually required, which prolongs the imaging time and affects patient comfort and scanner efficiency. 2. **Noise interference**: - Even when the very - low - field MRI scanner works in an electromagnetic shielding room, there are still Gaussian white noise and Johnson noise generated by the human body. - These noises will further reduce the image quality, especially when using the traditional high - field MRI parallel imaging method. 3. **Limitations of parallel imaging methods**: - Although parallel imaging (PI) techniques are mature in high - field MRI, they do not work well when directly applied to very - low - field MRI. - It is necessary to redesign the parallel imaging algorithm and phased - array coil according to the characteristics of very - low - field MRI. ### Solutions: 1. **Parallel imaging algorithm (KF - NL - GRAPPA)**: - **Kalman filter denoising**: Use the Kalman filter based on linear prediction to denoise the undersampled k - space data. - **Non - linear mapping**: Non - linearly map the denoised k - space data from the original space to a high - dimensional feature space. - **Estimate missing data**: Estimate the missing phase - encoding lines in the feature space through the acquired phase - encoding lines, and finally reconstruct the image. 2. **8 - channel phased - array coil design**: - **Geometric structure optimization**: According to the experimental results, a coil skeleton that closely fits the head is designed, balancing the signal - to - noise ratio performance and patient comfort. - **Decoupling technology**: Adopt geometric overlap and low - input - impedance (LII) pre - amplifier decoupling technology to reduce the coupling between coil elements and improve the signal - to - noise ratio and independence. ### Experimental verification: The paper verifies the performance of the proposed parallel imaging technique (KF - NL - GRAPPA) through head imaging experiments on healthy volunteers and compares it with the traditional GRAPPA and NL - GRAPPA methods. The results show that KF - NL - GRAPPA can obtain image quality comparable to or even better than that of fully - sampled images within half of the scanning time, and the signal - to - noise ratio of the reconstructed image exceeds that of the fully - sampled image. ### Summary: The purpose of this paper is to solve the problems of weak magnetic resonance signal, large noise interference, and long imaging time in very - low - field MRI by improving parallel imaging techniques and coil design, thereby improving the image quality and imaging efficiency of very - low - field MRI.