Improved balance of artifact / noise level and fine structure preservation in highly accelerated ppi

Dengrong Jiang,Kui Ying,Feng Huang
2012-01-01
Abstract:Target Audience Scientists and clinicians interested in highly accelerated MRI. Purpose Based on spatial sensitivity of coil arrays, Partially Parallel Imaging (PPI) [1] has been widely used in clinical practice to accelerate acquisition, but at the cost of reduced Signal-to-Noise Ratio (SNR). Often, regularization schemes are used to preserve SNR. However, existing regularization schemes have the difficulty to balance the reduction of artifact/noise level and the preservation of boundaries and fine structures, especially when the acceleration factor is high. The purpose of this work is to achieve low noise/artifact level with minimal impact on fine structures at high acceleration factors. Preliminary results demonstrated that lower root-mean-square-error (RMSE) and better preserved fine structures can be achieved by using the proposed method than CS-SENSE [2]. Low errors image was reconstructed at acceleration factor as high as 8 with an 8-channel head coil. Methods Patch-wise non-local sparse [3,4]: The transform domain determines the sparsity, and is the key to compressive sensing based denoising. In patch-wise non-local sparse, the transform domain is the clustered image fragments (patches) according to the similarity to exemplar patterns. In this work, we adopt low-rank as the sparsity constraint onto the transform domain to exploit the correlation among similar patches. Combination with PPI: To combine patch-wise non-local sparse with sensitivity encoding (SENSE), the following objective function was proposed:
What problem does this paper attempt to address?