A novel mechanical design of a bolometric array for the CROSS double-beta decay experiment

D. Auguste,A.S. Barabash,V. Berest,L. Bergé,J.M. Calvo-Mozota,P. Carniti,M. Chapellier,I. Dafinei,F.A. Danevich,T. Dixon,L. Dumoulin,F. Ferri,A. Gallas,A. Giuliani,C. Gotti,P. Gras,A. Ianni,L. Imbert,H. Khalife,V.V. Kobychev,S.I. Konovalov,P. Loaiza,P. de Marcillac,S. Marnieros,C.A. Marrache-Kikuchi,M. Martinez,C. Nones,E. Olivieri,A. Ortiz de Solórzano,Y. Peinaud,G. Pessina,D.V. Poda,Ph. Rosier,J.A. Scarpaci,V.I. Tretyak,V.I. Umatov,M.M. Zarytskyy,A. Zolotarova
2024-07-24
Abstract:The CROSS experiment will search for neutrinoless double-beta decay using a specific mechanical structure to hold thermal detectors. The design of the structure was tuned to minimize the background contribution, keeping an optimal detector performance. A single module of the structure holds two scintillating bolometers (with a crystal size of 45x45x45 mm and a Ge slab facing the crystal's upper side) in the Cu frame, allowing for a modular construction of a large-scale array. Two designs are released: the initial $Thick$ version contains around 15% of Cu over the crystal mass (lithium molybdate, LMO), while this ratio is reduced to ~6% in a finer ($Slim$) design. Both designs were tested extensively at aboveground (IJCLab, France) and underground (LSC, Spain) laboratories. In particular, at LSC we used a pulse-tube-based CROSS facility to operate a 6-crystal array of LMOs enriched/depleted in $^{100}$Mo. The tested LMOs show high spectrometric performance in both designs; notably, the measured energy resolution is 5--7 keV FWHM at 2615 keV $\gamma$s, nearby the Q-value of $^{100}$Mo (3034 keV). Due to the absence of a reflective cavity around LMOs, a low scintillation signal is detected by Ge bolometers: ~0.3 keV (150 photons) for 1-MeV $\gamma$($\beta$) LMO-event. Despite that, an acceptable separation between $\alpha$ and $\gamma$($\beta$) events is achieved with most devices. The highest efficiency is reached with light detectors in the $Thick$ design thanks to a lower baseline noise width (0.05--0.09 keV RMS) when compared to that obtained in the $Slim$ version (0.10--0.35 keV RMS). Given the pivotal role of bolometric photodetectors for particle identification and random coincidences rejection, we will use the structure here described with upgraded light detectors, featuring thermal signal amplification via the Neganov-Trofimov-Luke effect, as also demonstrated in the present work.
Instrumentation and Detectors
What problem does this paper attempt to address?
The paper primarily focuses on the study of a new mechanical structure used in the CROSS double beta decay experiment, especially the design used to support the thermal detector array. Specifically, the paper attempts to address the following key issues: 1. **Minimization of Background Radiation Contribution**: By optimizing the mechanical structure design to reduce the impact of background radiation, maintaining the optimal performance of the detector. 2. **Modular Construction of Large-Scale Arrays**: A single module capable of accommodating two scintillating thermal detectors (crystal size of 45×45×45 millimeters, with a germanium plate facing the upper side of the crystal) has been designed to achieve modular construction of large-scale arrays. 3. **Reduction of Copper to Crystal Mass Ratio**: Two design versions, Thick and Slim, have been proposed. The copper content in the Thick version is about 15% of the crystal mass, while in the Slim version, this ratio is reduced to about 6%, aiming to reduce background radiation sources. 4. **Optimization of Light Signal Detection**: Although no reflective cavity is used around the crystal, resulting in a lower scintillation signal detected by the germanium thermal detector (about 0.3 keV), an acceptable separation between α particle and γ/β particle events is still achieved. 5. **Improved Efficiency of Lightweight Detectors**: In the Thick design, due to the lower baseline noise width (0.05–0.09 keV RMS), higher efficiency is obtained compared to the Slim design (0.10–0.35 keV RMS). 6. **Application of Upgraded Lightweight Detectors**: To further improve particle identification capabilities and the exclusion of random coincident events, there are plans to use upgraded lightweight detectors that can amplify thermal signals through the Neganov-Trofimov-Luke effect. In summary, the goal of this research is to develop a new mechanical structure aimed at providing an efficient, low-background thermal detector array solution for the CROSS double beta decay experiment, enhancing the experiment's ability to detect rare nuclear transitions—neutrinoless double beta decay.