First test of a $\mathrm{CdMoO_4}$ scintillating bolometer for neutrinoless double beta decay experiments with $\mathrm{{}^{116}Cd}$ and $\mathrm{{}^{100}Mo}$ nuclides

M. Xue,D. V. Poda,Y. Zhang,H. Khalife,A. Giuliani,H. Peng,P. de Marcillac,E. Olivier,S. Wen,K. Zhao,Y. Wei,V. Novati,A. S. Zolotarova,S. Marnieros,C. Nones,T. Redon,Z. Xu,X. Wang,P. Chen,H. Chen,L. Dumoulin,D.V. Poda,A.S. Zolotarova
DOI: https://doi.org/10.48550/arXiv.1907.08039
2019-07-18
Instrumentation and Detectors
Abstract:A large cylindrical cadmium molybdate crystal with natural isotopic abundance has been used to fabricate a scintillating bolometer. The measurement was performed above ground at milli-Kelvin temperature, with simultaneous readout of the heat and the scintillation light. The energy resolution as FWHM has achieved from 5 keV (at 238 keV) to 13 keV (at 2615 keV). We present the results of the $\alpha$ versus $\beta$/$\gamma$ events discrimination. The low internal trace contamination of the $\mathrm{CdMoO_4}$ crystal was evaluated as well. The detector performance with preliminary positive indications proves that cadmium molybdate crystal is a promising absorber for neutrinoless double beta decay scintillating bolometric experiments with $\mathrm{{}^{116}Cd}$ and $\mathrm{{}^{100}Mo}$ nuclides in the next-generation technique.