First demonstration of a TES based cryogenic Li$_2$MoO$_4$detector for neutrinoless double beta decay search

G. Bratrud,C. L. Chang,R. Chen,E. Cudmore,E. Figueroa-Feliciano,Z. Hong,K. T. Kennard,S. Lewis,M. Lisovenko,L. O. Mateo,V. Novati,V. Novosad,E. Oliveri,R. Ren,J. A. Scarpaci,B. Schmidt,G. Wang,L. Winslow,V. G. Yefremenko,J. Zhang,D. Baxter,M. Hollister,C. James,P. Lukens,D. J. Temples
2024-06-04
Abstract:Cryogenic calorimetric experiments to search for neutrinoless double-beta decay ($0\nu\beta\beta$) are highly competitive, scalable and versatile in isotope. The largest planned detector array, CUPID, is comprised of about 1500 individual Li$_2^{100}$MoO$_{4}$ detector modules with a further scale up envisioned for a follow up experiment (CUPID-1T). In this article, we present a novel detector concept targeting this second stage with a low impedance TES based readout for the Li$_2$MoO$_{4}$ absorber that is easily mass-produced and lends itself to a multiplexed readout. We present the detector design and results from a first prototype detector operated at the NEXUS shallow underground facility at Fermilab. The detector is a 2-cm-side cube with 21$\,$g mass that is strongly thermally coupled to its readout chip to allow rise-times of $\sim$0.5$\,$ms. This design is more than one order of magnitude faster than present NTD based detectors and is hence expected to effectively mitigate backgrounds generated through the pile-up of two independent two neutrino decay events coinciding close in time. Together with a baseline resolution of 1.95$\,$keV (FWHM) these performance parameters extrapolate to a background index from pile-up as low as $5\cdot 10^{-6}\,$counts/keV/kg/yr in CUPID size crystals. The detector was calibrated up to the MeV region showing sufficient dynamic range for $0\nu\beta\beta$ searches. In combination with a SuperCDMS HVeV detector this setup also allowed us to perform a precision measurement of the scintillation time constants of Li$_2$MoO$_{4}$. The crystal showed a significant fast scintillation emission with O(10$\,\mu$s) time-scale, more than an order below the detector response of presently considered light detectors suggesting the possibility of further progress in pile-up rejection through better light detectors in the future.
High Energy Physics - Experiment,Nuclear Experiment,Instrumentation and Detectors
What problem does this paper attempt to address?
### Problems the Paper Attempts to Solve This paper aims to address the issue of background noise in the detection of neutrinoless double-beta decay (0νββ). Specifically, the research team has developed a new detector concept based on low-temperature thermoelectric sensors (TES) to improve the performance of lithium molybdate (Li₂MoO₄) crystal detectors. #### Background Issues 1. **Background Noise**: In existing low-temperature thermoelectric detectors, pile-up background noise caused by two independent double-beta decay events (2νββ) occurring close in time is a major factor affecting detection sensitivity. 2. **Response Speed**: The currently used neutron transmutation doped germanium thermistor (NTD) readout method has a slow response time (approximately 10 milliseconds), which is ineffective in suppressing this pile-up background noise. 3. **Energy Spectrum Resolution**: Improving energy spectrum resolution is crucial for reducing background noise and enhancing detection efficiency. #### Solutions 1. **New Detector Design**: The research team proposed a low-impedance TES readout method for the lithium molybdate (Li₂MoO₄) absorber. This design can achieve a faster response time (approximately 0.5 milliseconds), which is more than an order of magnitude faster than existing NTD detectors. 2. **Multiplexed Readout**: This new detector design also supports multiplexed readout, which helps reduce the thermal load on the low-temperature system. 3. **Performance Parameters**: Preliminary tests indicate that the detector has a baseline resolution of 1.95 keV (FWHM). It is expected that in CUPID-scale crystals, the pile-up background noise index can be reduced to 5 × 10⁻⁶ counts/keV/kg/yr. 4. **Dynamic Range**: The detector has been calibrated in the MeV region, showing sufficient dynamic range suitable for 0νββ searches. With these improvements, the research team hopes to significantly reduce background noise and enhance the sensitivity of neutrinoless double-beta decay detection in future CUPID-1T experiments.