A sinusoidal current generator and a TEB decomposer for measuring bioimpedance in a cardiac pacemaker, using an analog wavelet filter
Guilherme Tarcísio Leal,Gabriela Meira de Moura Rodrigues,Sandro Augusto Pavlik Haddad
DOI: https://doi.org/10.1007/s10470-021-01896-x
IF: 1.321
2021-06-13
Analog Integrated Circuits and Signal Processing
Abstract:This paper describes a novel sinusoidal signal generator for thoracic bioimpedance measurement system and signals decomposition filter, based on an analog wavelet filter approach. The proposed system can be adjusted in amplitude, frequency, and phase for differents sinusoidal injection currents, presenting a harmonic distortion rate lower than and a spurious-free dynamic range (SFDR) of . Applying the proposed sinusoidal generator to a bioimpedance measurement system (TEB), the obtained recovery accuracy was . Also, an analog wavelet filter approach was implemented on a new design of a TEB decomposing system, in order to extract its cardiac and respiratory components. From a system validation of the TEB decomposer, it was able to recover efficiently the cardiac and respiratory signals with a root mean square error (RMSE) below , even under dynamic variations of its parameters over time. Finally, previous works indicate that the proposed analog wavelet filter system can be implemented using a ultra low-power circuit design techniques, which allows its use in implantable systems such as pacemakers.
engineering, electrical & electronic,computer science, hardware & architecture