Thermalization of a SQUID chip at cryogenic temperature: Thermal conductance measurement for GE 7031 Varnish Glue, Apiezon N Grease and Rubber Cement between 20 and 200 mK

M. D'Andrea,G. Torrioli,C. Macculi,M. Kiviranta
DOI: https://doi.org/10.1007/s10909-023-03038-1
2024-01-22
Abstract:In the context of the ATHENA X-IFU Cryogenic AntiCoincidence Detector (CryoAC) development, we have studied the thermalization properties of a 2mm x 2mm SQUID chip. The chip is glued on a front-end PCB and operated on the cold stage of a dilution refrigerator (TBASE < 20 mK). We performed thermal conductance measurements by using different materials to glue the SQUID chip on the PCB. These have been repeated in subsequent cryostat runs, to highlight degradation effects due to thermal cycles. Here, we present the results obtained by glues and greases widely used in cryogenic environments, i.e. GE 7031 Varnish Glue, Apiezon N Grease and Rubber Cement.
Instrumentation and Methods for Astrophysics,Instrumentation and Detectors
What problem does this paper attempt to address?
The paper attempts to address the issue of how to select suitable adhesive materials to fix the SQUID chip (Superconducting Quantum Interference Device) onto the printed circuit board (PCB) in low-temperature environments to ensure good thermal performance. Specifically, the researchers focus on the impact of different types of adhesives (such as GE 7031 varnish, Apiezon N grease, and rubber cement) on the thermal conductivity between the SQUID chip and the PCB during the development of the ATHENA X-IFU cryogenic anti-coincidence detector (CryoAC). By measuring the thermal conductivity of these materials in the temperature range of 20 mK to 200 mK and evaluating their performance changes after multiple thermal cycles, the researchers hope to provide important reference data for the further development of CryoAC. This study is crucial for ensuring the stable operation of CryoAC under extreme low-temperature conditions and also provides valuable information for other applications in the field of cryogenic detection technology.