A study of the irradiation technique used for the external beam radiotherapy of retinoblastoma

L. Brualla,P.A. Mayorga,A. Flühs,A.M. Lallena,J. Sempau,W. Sauerwein
DOI: https://doi.org/10.1088/0031-9155/57/22/7741
2018-09-25
Abstract:Purpose: Retinoblastoma (RB) is the most common eye tumor in childhood and can be treated external radiotherapy. The purpose of this work is to evaluate the adequacy of Monte Carlo simulations and the accuracy of a commercial treatment planning system by means of experimental measurements. Dose measurements in water were performed using a dedicated collimator. Methods: A 6MV Varian Clinac 2100 C/D and a dedicated collimator are used for RB treatment. The collimator conforms a D-shaped off-axis field whose irradiated area can be either 5.2 or 3.1cm$^2$. Depth dose distributions and lateral profiles were measured and compared with Monte Carlo simulations run with PENELOPE and with calculations performed with the analytical anisotropic algorithm (AAA) using the gamma test. Results: PENELOPE simulations agree well with the experimental data with discrepancies in the dose profiles less than 3mm of distance-to-agreement and 3% of dose. Discrepancies between the results of AAA and the experimental data reach 3mm and 6%. The agreement in the penumbra region between AAA and the experiment is noticeably worse than that between the latter and PENELOPE. The percentage of voxels passing the gamma test when comparing PENELOPE (AAA) and the experiment is on average 99% (93%) assuming a 3mm distance-to-agreement and a discrepancy of 3% of dose. Conclusions: Although the discrepancies between AAA and experimental results are noticeable, it is possible to consider this algorithm for routine treatment planning of RB patients, provided the limitations of the algorithm are known and taken into account by the medical physicist. Monte Carlo simulation is essential for knowing these limitations. Monte Carlo simulation is required for optimizing the treatment technique and the dedicated collimator.
Medical Physics
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to evaluate and improve the techniques used for external beam radiotherapy of retinoblastoma. Specifically, the research aims to improve the current techniques by reducing the irradiated volume and minimizing the dose to the facial bones, so as to make better use of the advantages of external beam radiotherapy. To achieve this goal, the researchers used Monte Carlo simulations and a commercial treatment planning system (the Analytical Anisotropic Algorithm, AAA in the Eclipse treatment planning system) for dose measurement, and verified the accuracy of these methods through experimental data. The focus of the study is to evaluate the accuracy of these two calculation methods in predicting dose distribution, especially their ability to predict the dose distribution of small fields and off - axis fields, in order to optimize the treatment techniques and the design of dedicated collimators, reduce the radiation dose to healthy tissues, and thus reduce the risk of radiation - induced secondary malignancies.