Pre-treatment radiotherapy dose verification using Monte Carlo doselet modulation in a spherical phantom

Reid W Townson,Sergei Zavgorodni
DOI: https://doi.org/10.1088/0031-9155/59/8/1923
2013-10-01
Abstract:Due to the increasing complexity of radiotherapy delivery, accurate dose verification has become an essential part of the clinical treatment process. The purpose of this work was to develop an electronic portal image (EPI) based pre-treatment verification technique capable of quickly reconstructing 3D dose distributions from both coplanar and non-coplanar treatments. The dose reconstruction is performed in a spherical water phantom by modulating, based on EPID measurements, pre-calculated Monte Carlo (MC) doselets defined on a spherical coordinate system. This is called the spherical doselet modulation (SDM) method. This technique essentially eliminates the statistical uncertainty of the MC dose calculations by exploiting both azimuthal symmetry in a patient-independent phase-space and symmetry of a virtual spherical water phantom. The symmetry also allows the number of doselets necessary for dose reconstruction to be reduced by a factor of about 250. In this work, 51 doselets were used. The SDM method mitigates the most computationally intensive part of this type of dose reconstruction - reading, weighting and summing dose matrices. The accuracy of the system was tested against MC calculations as well as our previously reported phase-space modulation (PSM) method using a series of open field and IMRT cases. The mean chi- and gamma-test 3% / 3 mm success rates of the SDM method were 98.6% and 99.5%, respectively, when compared to full MC simulation. The total calculation time was 96 seconds per treatment field on a single processor core.
Medical Physics
What problem does this paper attempt to address?