Dry demagnetization cryostat for sub-millikelvin helium experiments: refrigeration and thermometry

I. Todoshchenko,J.-P. Kaikkonen,R. Blaauwgeers,P. J. Hakonen,A. Savin
DOI: https://doi.org/10.1063/1.4891619
2014-07-31
Abstract:We demonstrate successful "dry" refrigeration of quantum fluids down to $T=0.16$\,mK by using copper nuclear demagnetization stage that is pre-cooled by a pulse-tube-based dilution refrigerator. This type of refrigeration delivers a flexible and simple sub-mK solution to a variety of needs including experiments with superfluid $^3$He. Our central design principle was to eliminate relative vibrations between the high-field magnet and the nuclear refrigeration stage, which resulted in the minimum heat leak of $Q=4.4$\,nW obtained in field of 35\,mT. For thermometry, we employed a quartz tuning fork immersed into liquid $^3$He. We show that the fork oscillator can be considered as self-calibrating in superfluid $^3$He at the crossover point from hydrodynamic into ballistic quasiparticle regime.
Instrumentation and Detectors,Other Condensed Matter
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to achieve deep sub - millikelvin temperature cooling of quantum fluids (such as superfluid helium - 3) under "dry" (helium - free) conditions. Specifically, the authors demonstrated a "dry" refrigeration technique based on the copper - nuclear demagnetization stage, which is precooled by a pulse - tube dilution refrigerator and can reduce the temperature to 0.16 mK. The key to this technique lies in eliminating the relative vibration between the high - field magnet and the nuclear refrigeration stage, thereby reducing heat leakage. In addition, the paper also explored the method of using a quartz tuning fork immersed in liquid helium - 3 as a thermometer, especially its self - calibration characteristics at the transition point from the hydrodynamic to the ballistic quasiparticle mechanism. ### Core problems of the paper: 1. **Helium - free refrigeration at deep sub - millikelvin temperatures**: Through the copper - nuclear demagnetization technique combined with a pulse - tube dilution refrigerator, cooling of superfluid helium - 3 is achieved, reaching a low temperature of 0.16 mK. 2. **Reducing heat leakage**: By designing to eliminate the relative vibration between the high - field magnet and the nuclear refrigeration stage, heat leakage is minimized (4.4 nW). 3. **Temperature measurement method**: A temperature measurement method based on a quartz tuning - fork oscillator has been developed. This method has self - calibration characteristics in superfluid helium - 3 and can accurately measure temperature at extremely low temperatures. ### Main contributions: - **Technical implementation**: Successfully demonstrated the feasibility of achieving deep sub - millikelvin temperatures under helium - free conditions. - **Innovative design**: Significantly reduced heat leakage through mechanical fixation and vibration reduction. - **Temperature measurement**: Proposed a new temperature measurement method, using the self - calibration characteristics of the quartz tuning - fork oscillator to improve the accuracy of temperature measurement. ### Application prospects: - **Quantum experiments**: Provides a new platform for quantum experiments requiring extremely low temperatures. - **Scientific research**: Promotes in - depth research on superfluid helium - 3 and other quantum fluids. - **Industrial applications**: Makes it possible for helium - free refrigeration techniques to be applied in various fields (such as microscopy, imaging, medical, space applications, etc.). Through these innovations, the paper laid the foundation for the helium - free refrigeration technology at deep sub - millikelvin temperatures and provided important technical support for future scientific research and industrial applications.