A linear calibration method on DNL error for energy spectrum

Lu Bo,Fu Yan-Hong,Chen Yong,Cui Wei-Wei,Wang Yu-Sa,Xu Yu-Peng,Yin Guo-He,Yang Yan-Ji,Zhang Yi,Huo Jia,Li Wei,Cheng Tian-Xiang,Hu Wei,Han Da-Wei,Li Mao-Shun,Wang Juan,Wang Yu,Liu Xiao-Yang,Zhu Yue,Zhao Zhong-Yi,Zhang Ya
DOI: https://doi.org/10.48550/arXiv.1404.6303
2014-04-25
Abstract:A calibration method aimed for the differential nonlinearity (DNL) error of the Low Energy X-ray Instrument (LE) onboard the Hard X-ray Modulation Telescope (HXMT) is presented, which is independent with electronic systems used as testing platform and is only determined by the analog-to-digital converter (ADC) itself. Exploring this method, ADCs that are used within the flight model phase of HXMT-LE can be calibrated individually and independently by a non-destructive and low-cost way, greatly alleviating the complexity of the problem. As a result, the performance of the energy spectrum can be significantly improved, further more, noise reduced and resolution enhanced.
Instrumentation and Methods for Astrophysics,Instrumentation and Detectors
What problem does this paper attempt to address?