Beam test of a one-dimensional position sensitive chamber on synchrotron radiation

Liu Mei,Dai Hong-Liang,Qi Hui-Rong,Zhuang Bao-An,Zhang Jian,Liu Rong-Guang,Zhu Qi-Ming,Ouyang Qun,Chen Yuan-Bo,Jiang Xiao-Shan,Wang Ya-Jie,Liu Peng,Chang Guang-Cai
DOI: https://doi.org/10.48550/arXiv.1305.5897
2013-05-25
Abstract:One-dimensional single-wire chamber was developed to provide high position resolution for powder diffraction experiments with synchrotron radiation. A diffraction test using the sample of SiO2 has been accomplished at 1W2B laboratory of Beijing Synchrotron Radiation Source. The data of beam test were analyzed and some diffraction angles were obtained. The experimental results were in good agreement with standard data from ICDD powder diffraction file. The precision of diffraction angles was 1% to 4.7%. Most of relative errors between measured values of diffraction angles and existing data were less than 1%. As for the detector, the best position resolution in the test was 138 um (sigma value) with an X-ray tube. Finally, discussions of the results were given. The major factor that affected the precision of measurement was deviation from the flat structure of detector. The effect was analyzed and it came to a conclusion that it would be the optimal measurement scheme when the distance between the powder sample and detector was from 400 mm to 600 mm.
Instrumentation and Detectors
What problem does this paper attempt to address?