Detection of setup errors with a body‐surface laser‐scanning system for whole‐breast irradiation after breast‐conserving surgery

P. Jiang,Ziyi Liu,Weijuan Jiang,A. Qu,Haitao Sun,Junjie Wang
DOI: https://doi.org/10.1002/acm2.13578
2022-03-16
Journal of Applied Clinical Medical Physics
Abstract:Abstract Purpose We compared the setup errors determined by an optical imaging system (OSIS) in women who received breast‐conserving surgery (BCS) followed by whole‐breast radiotherapy (WBRT) with those from cone‐beam computed tomography (CBCT) carried out routinely. Methods We compared 130 setup errors in 10 patients undergoing WBRT following BCS by analyzing the translational and rotational couch shifts via CBCT and OSIS. Patients were treated with intensity‐modulated radiotherapy (IMRT). The patient outline extracted from the planning reference Computed tomography (CT) was used as the reference for OSIS and CBCT alignment during treatment. We detected the setup uncertainty using CBCT and OSIS at the first five fractionations of RT and then twice a week. Results The absolute translational setup error (mean ± Standard deviation (SD)) in x (lateral), y (longitudinal), and z (vertical) axes detected by the OSIS was 0.14 ± 0.18, 0.15 ± 0.14, and 0.13 ± 0.13 cm, respectively. The rotational setup error (mean ± SD) in Rx (pitch), Ry (roll), and Rz (yaw) axes was 0.77 ± 0.54, 0.76 ± 0.61, and 1.23 ± 0.95, respectively. Significant difference is observed only in one direction (Rx, p = 0.03) in the paired setup errors obtaining from OSIS and CBCT, without significant differences in five directions. Conclusion OSIS is a repeatable and reliable system that can be used to detect misalignments with accuracy, which is capable of supplementing CBCT for WBRT after BCS. We believe that an OSIS may be easier to use, quicker, and reduce overall dose as this method of patient alignment does not require ionizing radiation.
Medicine
What problem does this paper attempt to address?