Comparison of Home Sleep Devices and Sleep Study Testing in Hypoglossal Nerve Stimulation Patients

Phillip Huyett,Phoebe K. Yu,David Kent,Stephanie Stahl,Shalini Manchanda,Reena Mehra,Jessica Rundo,Kent Lee,Lauren Makey,Max Lundeen,Asim Roy
DOI: https://doi.org/10.1002/lary.31710
IF: 2.97
2024-08-22
The Laryngoscope
Abstract:Hypoglossal nerve stimulation is an effective treatment for obstructive sleep apnea. The initial adjustment period is based on subjective improvement such as improvement in snoring and daytime sleepiness. Home sleep technologies stand the potential to offer objective assessment of sleep apnea as patients make adjustments to therapy at home and assess readiness for a home sleep apnea test or in‐lab titration polysomnography. Objective Hypoglossal nerve stimulation (HGNS) is an implantable therapy for obstructive sleep apnea (OSA). Therapy efficacy is currently confirmed by a formal sleep study after empiric adjustment by the patient at home based on their subjective experience with the device. Home‐based longitudinal apnea hypopnea index (AHI) measurements have the potential to refine HGNS therapeutic amplitude selection with objective data. Our objective was to compare AHI derived from routine sleep studies and two different home sleep devices in new HGNS recipients. Methods Prospectively enrolled patients receiving HGNS therapy were provided a Sleep Tracking Mat (Withings, Issy‐les‐Moulineaux, France) and NightOwl peripheral arterial tonometry (PAT) sensor (Ectosense, Leuven, Belgium) for longitudinal, home AHI monitoring from 1 to 6 months post‐implant. Therapy efficacy was assessed at 3 and 6 months post‐implant using in‐lab polysomnography (PSG) or home sleep apnea test (HSAT). The sleep mat and PAT sensor AHI were compared against PSG and HSAT for accuracy of OSA severity identification. Results Sixty patients were enrolled across 5 centers and followed for 6 months. The sleep mat had sensitivity and specificity for identifying AHI <15 of 61% and 82% and AHI <30 of 77% and 100%. The PAT device had sensitivity and specificity for identifying AHI <15 of 57% and 77% and AHI <30 of 81% and 80%. Conclusions The sleep mat and PAT sensor demonstrated high sensitivity and specificity for detection of mild and moderate OSA in patients with HGNS therapy and may enable longitudinal objective monitoring of HGNS efficacy in the home setting. Level of Evidence 3 Laryngoscope, 2024
medicine, research & experimental,otorhinolaryngology
What problem does this paper attempt to address?