Dose measurements of simulated tumour movement during respiration using a polymer gel dosimeter

Aurimas Krauleidis,Todorka Dimitrova,Diana Adlienė
DOI: https://doi.org/10.1016/j.nimb.2024.165316
2024-03-09
Abstract:During radiotherapy treatment, there is always considerable risk that the timing of the treatment delivery and the volume movement results in degradation or undesired heterogeneities in the dose distribution. As there may be a simultaneous movement of the tumour, patient, and treatment machine, it is essential to minimize the uncertainties owing to motion in a clinical setting. Despite all clinical settings, tumour movement due to respiration is most important, because moving tumours can follow complex trajectories, that may affect patient treatment results, and cause difficulties in dose assessment. In this study, the effect of vertical movement on dose distribution was investigated. To overcome motion-related dosimetry challenges radiation-sensitive normoxic Methacrylic acid gels (nMAG) were used for dose evaluation. The selection of the gel dosimetry method was based on nMAG gel's ability to record corresponding target volume expansion via polymerized gel volume obtained during performed irradiation with simulated movement. Three identical sample batches of nMAG gels have been produced at once. Corresponding samples from each batch were irradiated to the same doses from the dose range up to 36 Gy and experimental results were averaged. A 6 MV FFF photon beam collimated to 1.0 cm 2 was used for irradiation. X-ray CT and NMR imaging procedures were used for the dose profile and dose sensitivity evaluation by applying a gamma passing rate criteria of 3 %/3 mm. The movement of the gel was responsible for the reduction of the peak maximum dose by approximately 13.6 %. The shape and uniformity of the dose profiles were changing when applying different irradiation scenarios. Dose sensitivity evaluated in dynamic mode using MRI-R2 technique was 0.025 s - 1 Gy - 1 . This study has shown that polymer gels are sensitive enough to detect distorted dose distributions and evaluate the doses, obtained due to the movement of the irradiated target.
physics, nuclear, atomic, molecular & chemical,nuclear science & technology,instruments & instrumentation
What problem does this paper attempt to address?