Evaluation of an x-ray CT polymer gel dosimetry system in the measurement of deformed dose

E Maynard,E Heath,M Hilts,A Jirasek
DOI: https://doi.org/10.1088/2057-1976/ab895a
2020-04-27
Abstract:This study is an evaluation of the use of a N-isopropylacrylamide (NIPAM)-based x-ray CT polymer geldosimetry (PGD) system in the measurement of deformed dose. This work also compares dose that ismeasured by the gel dosimetry system to dose calculated by a novel deformable dose accumulationalgorithm, defDOSXYZnrc, that uses direct voxel tracking. Deformable gels were first irradiatedusing a single 3.5 × 5 cm 2 open field and the static dose was compared to defDOSXYZnrc as a controlmeasurement. Gel measurement was found to be in excellent agreement with defDOSXYZnrc in the staticcase with gamma passing rates of 94.5% using a 3%/3 mm criterion and 93.3% using a 3%/2 mmcriterion. Following the static measurements, a deformable gel was irradiated with the same singlefield under an external compression of 25 mm and then released from this compression for dosimetricread out. The measured deformed dose was then compared to deformed dose calculated by defDOSXYZnrcbas...