Pulse oximeter bench tests under different simulated skin tones

Suvvi K. Narayana Swamy,Chenyang He,Barrie R. Hayes-Gill,Daniel J. Clark,Sarah Green,Stephen P. Morgan
DOI: https://doi.org/10.1007/s11517-024-03091-2
2024-04-24
Medical & Biological Engineering & Computing
Abstract:Abstract Pulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO 2 ) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO 2 values (~ 70 to 100%). Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx IITM, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO 2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO 2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO 2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO 2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical Abstract
engineering, biomedical,computer science, interdisciplinary applications,mathematical & computational biology,medical informatics
What problem does this paper attempt to address?