Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter

Shyqyri Haxha,Chike Nwibor,Mian Ali,Mohamed Sakel,Karen Saunders,Vladimir Dyo,Shakira Nabakooza
DOI: https://doi.org/10.3390/s24113301
IF: 3.9
2024-05-23
Sensors
Abstract:Pulse oximeters are widely used in hospitals and homes for measurement of blood oxygen saturation level (SpO2) and heart rate (HR). Concern has been raised regarding a possible bias in obtaining pulse oximeter measurements from different fingertips and the potential effect of skin pigmentation (white, brown, and dark). In this study, we obtained 600 SpO2 measurements from 20 volunteers using three UK NHS-approved commercial pulse oximeters alongside our custom-developed sensor, and used the Munsell colour system (5YR and 7.5YR cards) to classify the participants' skin pigmentation into three distinct categories (white, brown, and dark). The statistical analysis using ANOVA post hoc tests (Bonferroni correction), a Bland–Altman plot, and a correlation test were then carried out to determine if there was clinical significance in measuring the SpO2 from different fingertips and to highlight if skin pigmentation affects the accuracy of SpO2 measurement. The results indicate that although the three commercial pulse oximeters had different means and standard deviations, these differences had no clinical significance.
engineering, electrical & electronic,instruments & instrumentation,chemistry, analytical
What problem does this paper attempt to address?
The paper attempts to address the issue of the impact of skin pigmentation and finger selection on the accuracy of blood oxygen saturation (SpO₂) measurements using Internet of Things (IoT)-based pulse oximeters. Specifically, the study focuses on the clinical significance of SpO₂ measurement results for different skin colors (white, brown, and black) and different finger positions, and explores whether the differences between commercial pulse oximeters are statistically significant. The study collected data from 20 volunteers, using three commercially available pulse oximeters approved by the UK's National Health Service (NHS) and a custom-developed sensor for measurements. The Munsell color system was used to classify the skin color of the participants. Statistical analyses included ANOVA post-hoc tests (Bonferroni correction), Bland-Altman plots, and correlation tests to determine the clinical significance of SpO₂ measurements at different finger positions and to highlight the impact of skin color on the accuracy of SpO₂ measurements. The results showed that although the means and standard deviations of the three commercial pulse oximeters differed, these differences were not clinically significant.