Impact of epidermal melanin on peripheral pulse oximetry measurements

James Rice,Jalan Dixon,Ethan Renfrew,Denise Nemeth,Akiva Finkelstein,Yan Gernhofer,Sam Hinckley,Rafael Chaparro,Teresa Samson,Cristina Pizano,Bernard Hildebrand,Steven Mifflin
DOI: https://doi.org/10.1101/2024.09.17.24313822
2024-09-18
Abstract:: Pulse oximetry is used widely to assess oxygen saturation in nearly all healthcare institutions. : This study aimed to investigate the relationship between epidermal melanin concentrations, determined by colorimetry, and discrepancies between arterial blood saturation (SaO2), determined by arterial blood gas (ABG), and peripheral oxygen saturation (SpO2) determined by pulse oximetry. : This observational, correlational study was conducted over a one-week period in 2024 at the University of the Incarnate Word School of Osteopathic Medicine. Twenty-eight healthy adult volunteers (14 male; 14 female) with varying skin pigmentation were recruited via convenience sampling. Participants with conditions affecting oxygen saturation were excluded. : Participants underwent non-invasive pulse oximetry to measure peripheral oxygen saturation (SpO₂) and invasive arterial blood gas (ABG) analysis to measure arterial oxygen saturation (SaO₂). Melanin concentrations were measured using DSM III ColorMeter on the dorsal and palmar aspects of the hand. SpO₂ was measured using the Masimo MightyStat® Rx Fingertip Pulse Oximeter, and ABG samples were analyzed using the i-STAT® 1 blood analyzer. : The primary outcome was the difference between SpO₂ and SaO₂ across varying melanin concentrations, hypothesized before data collection. : Analysis revealed a significant difference between SaO2 and SpO2 with a moderately positive correlation with average epidermal melanin concentrations (r = 0.432; 95% CI, 0.07 to 0.693; P = 0.022) and a moderately inverse correlation with average luminance (r = -0.492; 95% CI,- 0.731 to -0.145; P = 0.008). A strong inverse relationship was found between melanin and luminance (r = -0.85; 95% CI -0.928 to -0.698; P <0.001). PO2 emerged as the leading predictor in two regression models, explaining 50.5% of the variance when combined with melanin. : Taken together, these data suggest that pulse oximetry overestimates oxygen saturation in individuals with comparatively high epidermal melanin concentrations. Future investigations that include a broader age range and participants with various comorbidities may provide a more comprehensive understanding of the impact of epidermal melanin on the accuracy of pulse oximetry readings.
Public and Global Health
What problem does this paper attempt to address?